Measurement Timeouts

Description

The primary use of measurement timeouts is to regain control of the test set's GPIB in cases where the bus
could potentially "hang."
The time normally required for a measurement to complete may vary greatly depending on the individual
measurement, its settings, it's multiple measurement count value, and so forth. Because of this, you may need
to set the timeout longer than the default for measurements where a large number of multiple measurements
are requested or where measurement triggers may be infrequent.
Be careful when setting a timeout that is shorter than the default. It is possible to specify a timeout that is so
short the measurement does not even have a chance to begin. Measurement timeouts should always be at least
several seconds long.
Timeout units default to S (seconds). The seconds suffix is an optional part of the command. If you want MS
(milliseconds), US (microseconds) or NS (nanoseconds), you must specify these units in the suffix.

Timeout Default Values

Table 2. List of Timeouts and Default Values
Measurement Function
Transmit Power
Power versus Time
Phase and Frequency Error
Output RF Spectrum
Fast Bit Error Rate
Bit Error Rate
Analog Audio
Decoded Audio
also know as Uplink Speech Level
IQ Tuning
Dynamic Power
Call Connected Timeout
Default
Default
Timeout
Timeout State
Time
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
OFF
10 seconds
NA
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_gen_timeout.fm
Integrity
Indicator
Value
2
2
2
2
2
2
2
2
2
2
Not
included
Measurement Timeouts
128

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5515aE5515bE1960a

Table of Contents