Agilent Technologies 8960 10 Series Reference Manual page 419

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:ORFSpectrum:SWITching:FREQuency:POINts?
Function
Queries the number of frequency offsets currently on during an ORFS due to switching
measurement. See
Query
Range: 0 to 8
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:ORFSpectrum:TIMeout:TIME
Function
Selects/queries the timeout value in seconds that will be used for output RF spectrum
measurements when the
optional, if no units are specified than units default to S.
Setting
Range: .1 to 999
Resolution: .1
Query
Range: .1 to 999
Resolution: .1
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:TIMEOUT:TIME 1" !Sets the timeout value to
SETup:ORFSpectrum:TIMeout[:STIMe]
Function
Sets/queries the timeout value in seconds that will be used for output RF spectrum
measurements and turns the timeout state on. The units (S|MS) are optional, if no units are
specified than units default to S.
Setting
Range: .1 to 999
Resolution: .1
Query
Range: .1 to 999
Resolution: .1
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:TIMEOUT:STIME 12" !Sets the timeout value to
"Output RF Spectrum Measurement Description" on page 75
"SETup:ORFSpectrum:TIMeout:STATe"
!15 seconds.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_orfspectrum.fm
is ON. The units (S|MS) are
!10 seconds and the state to on.
SETup:ORFSpectrum