AFGenerator
AFGenerator:VOLTage:AMPlitude
Function
Sets/queries the amplitude for the audio generator when the audio generator state is on. The
units (V|MV) are optional, if no units are specified then units default to V.
Setting
Range: 0 - 9 v pk.
Resolution:
• .5mv pk. <= 1v pk. output
• 5mv pk. > 1v pk. output
Query
Range: 0 - 9 v pk.
Resolution:
• .5mv pk. <= 1v pk. output
• 5mv pk. > 1v pk. output
*RST setting
zero volts
Programming Example
OUTPUT 714;"AFGENERATOR:VOLTAGE:AMPLITUDE 1.414" !Sets the audio generator output
AFGenerator:VOLTage:STATe
Function
Sets/queries the audio generator state
Setting
0 | OFF | 1 |ON
Query
0|1
*RST setting
0|off
Programming Example
OUTPUT 714;"AFGENERATOR:VOLTAGE:STATE ON" !Set the audio generator state to ON.