Agilent Technologies 8960 10 Series Reference Manual page 407

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:IQTuning:CONTinuous
Function
Selects/queries the trigger state for I/Q Tuning measurements.
Setting
Continuous trigger mode: 1|ON
Single trigger mode: 0|OFF
Query
1|0
*RST
1|ON
Programming Example
OUTPUT 714;"SETUP:IQTUNING:CONTINUOUS OFF" !Sets trigger mode to single for
an I/Q Tuning measurement.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:IQTuning:COUNt:STATe
Function
Sets/queries the I/Q Tuning multi-measurement count state.
Setting
Range: 0|OFF|1|ON
Query
0|1
*RST
0|OFF
Programming Example
OUTPUT 714;"SETUP:IQTUNING:COUNT:STATE ON" !Turns on the multi-measurement
mode for I/Q Tuning measurements.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:IQTuning:COUNt:NUMBer
Function
Sets/queries the number of I/Q Tuning multi-measurements the test set makes when the
multi-measurement count state is on. See
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST
10
Programming Example
OUTPUT 714;"SETUP:IQTUNING:COUNT:NUMBER 80" !Sets the multi-measurement
count number for I/Q Tuning measurements to 80.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
"I/Q Tuning Measurement Description" on page
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_iq.fm
SETup:IQTuning
63.
407