Agilent Technologies 8960 10 Series Reference Manual page 442

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

STATus:OPERation
STATus:OPERation Condition Register Bit Assignment
The OPERation status register set contains bits which give an indication of conditions that are part of the test
set's normal operation.
Bit
Binary
Number
Weighting
15
32768
14
16384
13
8192
12
4096
11
2048
10
1024
9
512
8
256
7
128
6
64
5
32
4
16
3
8
2
4
1
2
0
1
442
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_status_operation.fm
Condition
Not Used. Defined by SCPI.
Reserved for future use.
Reserved for future use.
Processing
SYSTem:SYNChronized
Command
Reserved for future use
CALL Summary
NMRReady (New Measurement
Result Ready) Summary
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Extension Bit
Description
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This condition bit will be "pulsed" by the
SYSTem:SYNChronized command. This will allow
the status system to indicate that the input buffer is
synchronized to the point where this command is
parsed and that all prior sequential commands are
completed and all prior overlapped commands have
started.
This bit will always be 0.
This bit is the summary bit for the
OPERation:CALL register.
This bit is the summary bit for the
OPERation:NMRReady register.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.

Hide quick links: