I/Q Tuning Measurement Description
Single or Multi Measurements
The I/Q Tuning measurement can return either single or averaged measurement results.
• If you set the multi-measurement state OFF then only a single measurement is made at each offset.
• If you set the multi-measurement state ON, and the multi-measurement count number to a value greater
than one, then multiple measurements are made at each offset. The returned results are an average of
these measurements.
Types of Signals I/Q Tuning can Measure
I/Q Tuning measurements can be made on these types of input signals.
• Normal GSM TCH burst without a midamble.
• CW signal.
I/Q Tuning Input Signal Requirements
The I/Q Tuning measurement will complete and meet its measurement accuracy specifications under the
following input signal conditions.
• Input signal level is between -15 dBm and +43 dBm.
• Input signal level is within +3 dB and -10 dB of the expected input level.
• Signal must be within 500 kHz of expected frequency for RF Rise triggering to function.
Trigger Source
The trigger source depends on the type of input signal.
Recommended Trigger Source Settings
Input Signal Type
Normal GSM TCH burst without a
midamble
CW signal
Related Topics
*******************************************************
"Programming an I/Q Tuning Measurement" on page 65
"Test Adherence to Standards" on page 110
*******************************************************
64
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\meas_iq_desc.fm
Recommended Trigger Source
RF Rise
Immediate