AFGenerator:PULSe[:STATe]
Function
Sets/queries the audio generator pulse state.
The pulse state must be on when the test set's audio generator is used for audio stimulation
during a decoded audio measurement.
When the state is on, the audio signal from the test set is pulsed at a 10 Hz rate with a 50% duty
cycle.
The amplitude and frequency of the pulse is set with afgenerator commands. See
on page
Setting
Range: 0 | OFF | 1 |ON
Query
Range: 0|1
*RST setting
0|off
Programming Example
OUTPUT 714;"AFGENERATOR:PULSE ON" !Sets the audio generator pulse to ON.
AFGenerator:VOLTage[:SAMPlitude]
Function
Sets /queries the amplitude of the audio generator in volts and turns the state to on. The units
(V|MV) are optional, if no units are specified then units default to V.
Setting
Range: 0 - 9 v pk.
Resolution:
• .5mv pk. <= 1v pk. output
• 5mv pk. > 1v pk. output
Query
Range: 0 - 9 v pk.
Resolution:
• .5mv pk. <= 1v pk. output
• 5mv pk. > 1v pk. output
*RST setting
zero volts
Programming Example
OUTPUT 714;"AFGENERATOR:VOLTAGE 2.1" !Sets the state to on and the output
220.
!voltage to 2.1 volts.