Agilent Technologies 8960 10 Series Reference Manual page 345

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

FETCh:PVTime:POWer:TIME[OFFSet]:MINimum?
Function
Queries the minimum power levels, from a number of multi-measurements, at each user-settable
time offset appended to this command. Specified time values must correspond to user-settable
time offsets that are currently turned on, and must be rounded to the same values. (9.91 E+37
(NAN) will be returned for specified offsets that do not correspond to offsets currently turned on).
Power levels are relative to the power versus time carrier power measurement.
The
"SETup:PVTime:TIME[:OFFSet]"
and their time values. Measurements will be returned by this query in the same order they are
listed in the command.
Query
Power levels:
• Range: −100 dBc to +10 dBc and 9.91 E+37 (NAN)
• Resolution: 0.01 dB
Time offsets:
• Range: Up to 12 time offset values, corresponding to entries in the Power vs Time table of
user-defined time offsets currently turned on. The default units are s (seconds).
• Resolution: Rounded to the same value as displayed in the Power vs Time table and returned
by the
Programming Example
OUTPUT 714;"FETCH:PVTIME:POWER:TIME:OFFSET:MINIMUM? 0 US, 570.8 US"
!Returns the !minimum of power versus time measurements at the 0.0 ms and
!570.8 ms offsets.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
"SETup:PVTime:TIME[:OFFSet]"
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_fetch_pvtime.fm
command sets up the number of offsets that are turned on
query.
FETCh:PVTime
345