Invalid Measurement Results; Description - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

Invalid Measurement Results

Description

Invalid measurement results are returned by the test set when conditions such as signal level are not within
the present measurement range. Three different invalid measurement results are provided in order to help the
user understand the condition that caused the invalid result.
9.9E+37 = INFinity (Infinity)
-9.9E+37 = NINF (Negative Infinity)
9.91E+37 = NAN (Not A Number)
9.9E+37 (INFinity)
9.9E+37 is returned by the test set when, the measurement is out of range, results are far above the present
measurement range.
-9.9E+37 (NINFinity)
-9.91E+37 is returned by the test set when, the measurement is out of range, results are far below the present
measurement range.
9.91E+37 (NAN)
9.91E+37 is returned by the test set when, the measurement is out of range but it can not be determined if
measurement results are far above, or far below the measurement range.
If a measurement exceeds its measurement timeout value before a valid result is determined, 9.91E+37 is
returned.
FETCH? and READ? Invalid Results
When a FETCH? or READ? query is performed on a measurement with invalid results, the integrity indicator
will return a value of 1, indicating No Result Available.
Manual Users Invalid Results
Manual users will generally see four dashes , "----" on the test set display. When the measurement timeout
value has been exceeded, "Measurement Timeout" is displayed as well as the four dashes .
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_invalid_results.fm
Invalid Measurement Results
130