Returned Values - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

Programming an I/Q Tuning Measurement
320
UNTIL Meas_complete$="IQT"
330
OUTPUT 714;"FETCH:IQTUNING:ALL?"!Fetches the measurement integrity
340
!value and the relative power levels at the offset frequencies.
350
ENTER 714;Integrity,N270,N203,N135,N67,Carrier,P67,P135,P203,P270,Sr
360
PRINT "I/Q Tuning Measurement Results"
370
PRINT "Integrity = ";Integrity
380
PRINT "Spur Power = ";Sr
390
PRINT "Offset (kHz)
400
PRINT "------------
410
PRINT "-270.334
420
PRINT "-203.125
430
PRINT "-135.417
440
PRINT "-67.708
450
PRINT "0.000
460
PRINT "+67.708
470
PRINT "+135.417
480
PRINT "+203.125
490
PRINT "+270.334
500
END

Returned Values

The measurements returned by this program are:
• Integrity returns the measurement
with no errors).
• The signal level of the following offsets are measured relative to the signal level at the reference offset
(either Fc + 67.7083 kHz for all 0s or Fc -67.7083 kHz for all 1s). Note, if the TX I/Q Tuning measurement
multi-measurement command is set to ON the average of all the individual results at each offset are
returned.
— -270.833 kHz
— -203.125 kHz
— -135.417 kHz
— -67.7083 kHz
— Carrier Frequency
— +67.7083 kHz
— +135.417 kHz
— +203.125 kHz
— +270.833 kHz
66
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_iq_proc.fm
Level (dB)"
----------"
";N270
";N203
";N135
";N67
";Carrier
";P67
";P135
";P203
";P270
"Integrity Indicator" on page 125
(0 means a successful measurement

Hide quick links: