Agilent Technologies 8960 10 Series Reference Manual page 431

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:PVTime:TIMeout:STATe
Function
Sets/queries power versus time timeout state.
Setting
0 |OFF | 1 | ON
Query
0 | 1
*RST Setting
0 | OFF
Programming Example
OUTPUT 714;"SETUP:PVTIME:TIMEOUT:STATE ON" !Sets timeout state to on.
SETup:PVTime:TRIGger:DELay
Function
Selects/queries the trigger delay in seconds for power versus time measurements. The units
(S|MS|US|NS) are optional, if no units are specified then units default to S.
Range: −2.31 ms to +2.31 ms
Range
Resolution: 5 significant digits or 100 ns, whichever is greater
Range: −2.31 ms to +2.31 ms
Query
Resolution: 5 significant digits or 100 ns, whichever is greater
*RST Setting
0 seconds
Programming Example
OUTPUT 714;"SETUP:PVTIME:TRIGGER:DELAY 1.1MS" !Sets trigger delay value to 1.1
SETup:PVTime:TRIGger:SOURce
Function
Selects/queries the trigger source for power versus time measurements. See
Measurements" on page 149
Setting
AUTO|PROTocol|RISE|IMMediate
Query
AUTO|PROT|RISE|IMM
*RST Setting
AUTO
Programming Example
OUTPUT 714;"SETUP:PVTIME:TRIGGER:SOURCE AUTO" !Sets trigger source to AUT.
!milli-seconds.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pvtime.fm
SETup:PVTime
"Triggering of