Power versus Time Measurement Description

Trigger Source

Triggering choices available for the PvT measurement are RF rise, protocol, immediate, and auto. In most
cases, auto triggering provides the optimum measurement triggering condition for the PvT measurement.
When auto triggering is selected, the test set chooses a trigger source based on the optimum trigger source
available. For example, PvT measurements will automatically be triggered by a protocol trigger if a call is
connected or call processing events provide the protocol trigger source.
In situations where no protocol trigger is available, the test set will choose RF rise triggering for the PvT
measurement. An example of this situation might be when the test set is in test mode operating mode.
Table 5. Recommended Trigger Source Settings
Input Signal Type
Normal GSM TCH burst with mobile
station in active cell mode
RACH burst with mobile station in active
cell mode
Normal GSM TCH burst with mobile
station in test mode
RACH burst with mobile station in test
mode
Bursted signal with GMSK modulation but
no valid midamble
CW signal
For more information on measurement triggering, refer to
90
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\meas_pvt_desc.fm
Recommended Trigger Source
RF Rise or Protocol
RF Rise or Protocol
RF Rise
RF Rise
RF Rise
Immediate
"Triggering of Measurements" on page
149.