Agilent Technologies 8960 10 Series Reference Manual page 422

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:PFERror
SETup:PFERror:BSYNc
Function
Sets/queries the burst synchronization mode for phase/frequency measurements. See "Burst
Synchronization of Measurements" on page 115.
Setting
Range: MIDamble|AMPLitude|NONE
Query
Range: MID | AMPL | NONE
*RST Setting
MIDamble
Programming Example
OUTPUT 714;"SETUP:PFERROR:BSYNC MIDAMBLE" !Sets the burst synchronization.
SETup:PFERror:CONTinuous
Function
Sets/queries the trigger state for phase/frequency measurements.
Setting
Single trigger mode = 0|OFF
Continuous trigger mode = 1|ON
Query
0|1
*RST Setting
0|off
Programming Example
OUTPUT 714;"SETUP:PFERROR: CONTINUOUS OFF" !Specifies single trigger mode for
SETup:PFERror:COUNt[:SNUMber]
Function
Sets/queries the number of phase/frequency measurements the test set will make and turns the
multi-measurement count state on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
0 | off
Programming Example
OUTPUT 714;"SETUP:PFERROR:COUNT:SNUMBER 100" !Sets the value to 100 and the
422
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pferror.fm
!phase/frequency measurements.
!state to on