This product is a Safety Class I instrument (provided with a protective earthing
WARNING
ground incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the product is likely to make the product
dangerous. Intentional interruption is prohibited.
Whenever it is likely that the protection has been impaired, the instrument must be
made inoperative and be secured against any unintended operation.
If this instrument is to be energized via an autotransformer (for voltage reduction),
make sure the common terminal is connected to the earth terminal of the power
source.
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used in a normal condition (in which all
means for protection are intact) only.
No operator serviceable parts in this product. Refer servicing to qualified personnel.
To prevent electrical shock, do not remove covers.
Servicing instructions are for use by qualified personnel only. To avoid electrical
shock, do not perform any servicing unless you are qualified to do so.
The opening of covers or removal of parts is likely to expose dangerous voltages.
Disconnect the product from all voltage sources while it is being opened.
The power cord is connected to internal capacitors that my remain live for
5 seconds after disconnecting the plug from its power supply.
For Continued protection against fire hazard, replace the line fuse(s) only with 250 V
fuse(s) or the same current rating and type (for example, normal blow or time delay).
Do not use repaired fuses or short circuited fuseholders.
Always use the three-prong ac power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause product damage.
This product is designed for use in Installation Category II and Pollution Degree 2 per
IEC 1010 and IEC 664 respectively. FOR INDOOR USE ONLY.
This product has autoranging line voltage input, be sure the supply voltage is within
the specified range.
To prevent electrical shock, disconnect instrument from mains (line) before cleaning.
Use a dry cloth or one slightly dampened with water to clean the external case parts.
Do not attempt to clean internally.
Ventilation Requirements: When installing the product in a cabinet, the convection
into and out of the product must not be restricted. The ambient temperature (outside
the cabinet) must be less than the maximum operating temperature of the product by
4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
5