Input Signal Requirements; Trigger Source; Related Topics - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

Input Signal Requirements

The TX Power measurement will complete and meet its accuracy specification of less than ±0.6dB when.
• Level is between −20dBm and +3 dBm.
• Level within ±3 dB of the expected input level.
• Frequency is within ±100kHz of expected input frequency.

Trigger Source

Auto triggering is the recommended trigger source for each measurement allowing the test set may choose the
preferred trigger source. However the user may want to select the trigger source. See Table 8. on page 107
Table 8. Recommended Trigger Source Settings
Input Signal Type
Normal GSM TCH burst with mobile
station in active cell mode
RACH burst with mobile station in active
cell mode
Normal GSM TCH burst with mobile
station in test mode
RACH burst with mobile station in test
mode
Bursted signal with GMSK modulation but
no valid midamble
CW signal

Related Topics

*******************************************************
"Programming a Transmit Power Measurement" on page 108
"Test Adherence to Standards" on page 110
*******************************************************
Recommend Trigger Source
Amplitude or Protocol
Amplitude or Protocol
Amplitude
Amplitude
Amplitude
Immediate
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\meas_txp_desc.fm
Transmit Power Measurement Description
107

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5515aE5515bE1960a

Table of Contents