Triggering of Measurements
nanosecond per measurement and the units are in seconds. A negative value indicates the sampling should
occur prior to the trigger. The default is zero seconds which is preferred for most measurements.
Trigger Qualifier Description
When the trigger qualifier is on, the test set analyzes (samples) the input signal when a trigger is received. It
then determines if the input signal was valid by looking at its power level. If the power level during sampling
did not meet the requirements of a valid signal, the state returns to wait for trigger without making a
measurement. Trigger qualifier is available for TX Power and Phase Frequency Error measurements only.
If a valid signal is present, then it is qualified, and the samples are processed.
Related Topics
*******************************************************
"Integrity Indicator" on page 125
"SETup:ORFSpectrum:TRIGger:SOURce" on page 420
"SETup:PFERror:TRIGer:SOURce" on page 425
"SETup:PVTime:TRIGger:SOURce" on page 431
"SETup:IQTuning:TRIGger:SOURce" on page 411
"SETup:TXPower:TRIGger:SOURce" on page 435
*******************************************************
152