Agilent Technologies 8960 10 Series Reference Manual page 264

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

CALL:MS
CALL:MS:TXLevel:PCS
Function
Selects the MS uplink power control level for the PCS band. See
Parameters" on page
Additional commands can be appended to aid in controller/Mobile Station synchronization.
See
Setting
range: 0 to 31 (default 10)
resolution: 1
Query
range: 0 to 31
resolution: 1
*RST Setting
Band: PGSM
TXLevel: 15
Programming Example
OUTPUT 714;"CALL;MS;TXLEVEL:PCS 31"
CALL:MS:TXLevel:PGSM
Function
Selects the MS uplink power control level for the PGSM band. See
Parameters" on page
Additional commands can be appended to aid in controller/Mobile Station synchronization. See
"Call Processing Subsystem Overlapped Commands" on page
Setting
range: 0 to 31 (default 15)
resolution: 1
Query
range: 0 to 31
resolution: 1
*RST Setting
Band: PGSM
TXLevel: 15
Programming Example
OUTPUT 714;"CALL:MS:TXLEVEL:PGSM 22"
264
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_ms.fm
501.
"Call Processing Subsystem Overlapped Commands" on page
501.
"Frequency Banded
33.
"Frequency Banded
33.