Agilent Technologies 8960 10 Series Reference Manual page 433

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:TXPower:CONTinuous
Function
Sets/queries the trigger state for TX carrier power measurements.
Setting
Single trigger mode = 0|OFF Continuous trigger mode = 1|ON
Query
0|1
*RST Setting
0|off
Program
OUTPUT 714;"SETUP:TXPOWER: CONTINUOUS OFF"
Example
! specifies single trigger mode for TX Carrier Power measurements.
Programming Example
OUTPUT 714;"SETUP:TXPOWER: CONTINUOUS OFF" !Specifies single trigger mode for
SETup:TXPower:COUNt[:SNUMber]
Function
Sets/queries the number of TX carrier power measurements the test set will make and turns the
multi-measurement count state on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:TXPOWER:COUNT:SNUMBER 99" !Sets the state to on and the
SETup:TXPower:COUNt:NUMBer
Function
Sets/queries the number of TX power measurements the test set will make when the
multi-measurement state is on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:TXPOWER:COUNT:NUMBER 5" !Sets the TX Power multi-measurement
!TX Carrier Power measurements.
!multi-measurement count value
!to 99.
!count number to 5.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_txpower.fm
SETup:TXPower
433