SETup Subsystem
Description
The SETup subsystem is used to configure the test set for each measurement. Typical settings include:
• Specifying whether a measurement will run continuously or need to be INITiated.
• How a measurement is triggered
• How many measurements will be made each time a measurement is INITiated
Trigger arming for each measurement is controlled in the SETup subsystem. The choices are
NOTE
single or continuous. In most cases, it is a best practice (during remote operation) to use "single"
measurement mode. This simplifies the tasks of starting concurrent measurements, using the
INIT subsystem commands to determine which measurements are ready to be fetched, then
using the FETCh subsystem to obtain results. The command "SETup:CONTinuous:OFF sets all
measurements to "single" trigger mode.
Syntax Diagrams and Command Descriptions
"SETup:CONTinuous" on page 397
"SETup:AAUDio" on page 380
"SETup:BERRor" on page 385
"SETup:DAUDio" on page 398
"SETup:DPOWer" on page 403
"SETup:FBERror" on page 391
"SETup:IQTuning" on page 406
"SETup:ORFSpectrum" on page 412
"SETup:PFERror" on page 421
"SETup:PVTime" on page 426
"SETup:TXPower" on page 432