SETup:BERRor
SETup:BERRor:COUNt
Function
Sets/queries the number of BER measurements the test set will make when the count state is on
Setting
Range: 1 to 999,000
Resolution: 1
Query
Range: 1 to 999,000
Resolution: 1
*RST Setting
10,000
Programming Example
OUTPUT 714;"SETUP:BERROR:COUNT 880" !Set BER multi-measurement count
SETup:BERRor:LDControl:AUTO
Function
Sets/queries speech frames delay control mode. If speech frames delay control mode is automatic
(Auto), the test set will determine the frame delay value that will allow correlation between
uplink information bits with downlink information bits.
If speech frames delay control mode is manual (not Auto), the test set will use the frame delay
value entered in the Speech Frames Delay field. Refer to.
on page 389
Refer also to the
and its use in the BER measurement.
Setting
Range: 0 | OFF | 1 | ON
Query
0 | 1
*RST Setting
1 | auto
Related Topics
"SETup:BERRor:MANual:DELay" on page 389
Programming Example
OUTPUT 714;"SETUP:BERROR:LDCONTROL:AUTO OFF" !sets BER delay to manual the user
388
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_berror.fm
!to 880.
"Bit Error Measurement Description" on page 48
see"SETup:BERRor:MANual:DELay"
!must select the manual frame
!delay number.
for a description of frame delay