ORFS Troubleshooting

Interpreting Integrity Indicator values

See
"Integrity Indicator" on page
125.
If over range (5) is returned the input signal is likely to clip during the useful part of the burst or the ORFS TX
Power measurement has detected an over range.
If signal too noisy (10) is returned, the actual power at certain offsets is > 8 dB off from the expected level.
If under range (6) is returned; the measured power result is more than 10 dB below the expected input power
level. Under range is also indicated if, the input power is more than 2 dB below the calibrated range of the test
set's power detector for the RF Range setting. RF Range is automatically set based on the input power setting.
Input power is a combination of amplitude offset and expected power settings. See
"Receiver example" on page
564.