Agilent Technologies 8960 10 Series Reference Manual page 414

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:ORFSpectrum
SETup:ORFSpectrum:ICOunt:MAXimum?
Function
Queries the total number measurements made each time an ORFS measurement is initiated.
This number will vary depending on the number of offsets and number of multi-measurements
the user chooses.
The total number of measurements is calculated using the following formula:
ICO MAX = 1 + M + S
Where:
See
"Output RF Spectrum Measurement Description" on page 75
and switching offsets.
Query
Range: 1 to 29971
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:ORFSpectrum:MODulation:COUNt[:SNUMber]
Function
Sets/queries the number of output RF spectrum due to modulation multi-measurements the test
set will make. This command sets the count state to ON.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
20
Programming Example
OUTPUT 714"SETUP:ORFSPECTRUM:MODULATION:COUNT:SNUMBER 99" !Sets the value to 99
414
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_orfspectrum.fm
M= (the number of modulation offsets) × (the number of multi-measurements for ORFS due
to modulation).
S= (the number of switching offsets) × (the number of multi-measurements for ORFS due to
switching) .
for a description of modulation
!multi-measurements
!and the state to on.