Step 4: Establish an Active Link with Mobile Station

Process for Making a Mobile Station Originated Call

The recommended process for making a mobile station originated call is shown in
"Step 4: Figure 2. Process For Making A Mobile Station Originated Call" on page
180.
There is no facility in the test set to initiate a call connect from the mobile station. This must be accomplished
manually or through a test bus built into the mobile station.
If the call origination process fails it is necessary to send the CALL:END command to the test set to force
immediate termination of all processes associated with the current call origination. This ensures that if the
mobile station attempts another originate before all processes associated with the failed call origination have
been terminated, it will not be ignored. Note that if the test set is currently executing a call origination and it
receives another call origination command it will be ignored (that is, you are telling the test set to do
something it is already doing and hence it will accept the command but it will be ignored).
For mobile station originated calls where the call is originated by physically dialing a number (as opposed to
using a test bus) ensure that the call-state-change-detector time-out time is long enough to allow a human to
dial the number.