SETup:TXPower:TIMeout:STATe
Function
Selects/queries TX carrier power measurement timeout state.
Setting
0|OFF | 1|ON
Query
0|1
*RST Setting
0|off
Programming Example
OUTPUT 714;"SETUP:TXPOWER:COUNT:STATE 1" !Turns the TX carrier power timeout
SETup:TXPower:TRIGger:DELay
Function
Sets/queries the trigger delay in seconds for TX carrier power measurements. The units
(S|MS|US|NS) are optional, if no units are specified then units default to S.
Range: −2.31 ms to +2.31 ms
Setting
Resolution: 5 significant digits or 100 ns, whichever is greater
Range: −2.31 ms to +2.31 ms
Query
Resolution: 5 significant digits or 100 ns, whichever is greater
*RST Setting
zero seconds
Programming Example
OUTPUT 714;"SETUP:TXPOWER:TRIGGER:DELAY 1.5MS" !Set trigger delay time to
SETup:TXPower:TRIGger:SOURce
Function
Selects/queries the trigger source for TX carrier power measurements. See
Measurements" on page 149
Setting
AUTO|PROTocol|RISE|IMMediate
Query
AUTO|PROT|RISE|IMM
*RST Setting
AUTO
Programming Example
OUTPUT 714;"SETUP:TXPOWER:TRIGGER:SOURCE AUTO" !Sets trigger source to AUTO.
!state on.
!1.5 milliseconds
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_txpower.fm
SETup:TXPower
"Triggering of
435