CALL:RFGenerator
CALL[:CELL]:RFGenerator:FREQuency
Function
Sets/queries the RF Gen Frequency selection.
Operating mode = Test Mode and Downlink Function = CW. See
Behavior" on page
The units (Hz|KHz|MHz|GHz) are optional, if no units are specified then units default to
Hz.
Setting
range: 292 MHz to 2700 MHz
resolution: 1
Query
range: 292 MHz to 2700 MHz
resolution: 1
*RST Setting
939 MHz
Programming Example
OUTPUT 714;"CALL:CELL:RFGENERATOR:FREQUENCY 896.2MHZ"
CALL[:CELL]:RFGenerator:POWer[:SAMPLitude]
Function
Sets/queries the value for RF Gen Power and turns the state to ON.
The suffix dBm is optional.
Operating mode = Test Mode and Downlink Function = CW. See
Behavior" on page
The RF Gen Power field is affected when there is an amplitude offset, see
Related Configuration" on page
Setting
range: -10 to -100 dBm
resolution: .01 dBm
Query
range: -10 to -100 dBm
resolution: .01 dBm
*RST Setting
-85 dBm
Programming Example
OUTPUT 714;"CALL:CELL:RFGENERATOR:POWER:SAMPLITUDE -50DBM"
280
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_rfgenerator.fm
531.
531.
563.
"CW Test Function
"CW Test Function
"Measurement