Agilent Technologies 8960 10 Series Reference Manual page 277

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

CALL[:CELL]:RFGenerator:CHANnel[:SELected]
Function
Sets/queries the RF Gen Channel for the band already selected.
Operating mode = Test Mode and Downlink Function = CW. See
page
Additional commands can be appended to aid in controller/Mobile Station synchronization. See
"Call Processing Subsystem Overlapped Commands" on page 33
Setting
range:
• DCS channels 512 to 885
• EGSM channels 975 to 1023 and 0 to 124
• PCS channels 512 to 810
• PGSM channels 1 to 124
Query
range:
• DCS channels 512 to 885
• EGSM channels 975 to 1023 and 0 to 124
• PCS channels 512 to 810
• PGSM channels 1 to 124
*RST Setting
20 (PGSM band)
Programming Example
OUTPUT 714;"CALL:CELL:RFGENERATOR:CHANNEL:SELECTED 512" !Sets the RF Generator
531.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_rfgenerator.fm
"CW Test Function Behavior" on
for examples.
!channel to 512 for the
!band already selected.
CALL:RFGenerator
277

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5515aE5515bE1960a

Table of Contents