Agilent Technologies 8960 10 Series Reference Manual page 430

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:PVTime
SETup:PVTime:TIME:POINts?
Function
Queries the number of Measurement Offset points that are turned on during a power versus time
measurement.
Query
Range: 0 to 12
Resolution: 1
*RST Setting
12
Comments
This command is useful for determining how many time values will be returned in a
comma-separated list when the
power values will be returned when the
343
command is sent.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:PVTime:TIMeout[:STIMe]
Function
Sets/queries the timeout value in seconds that will be used for power versus time measurements.
This command also sets the timeout state to on. The units (S|MS) are optional, if no units are
specified then units default to S.
Setting
Range: .1 to 999
Resolution: .1
Query
Range: .1 to 999
Resolution: .1
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:PVTIME:TIMEOUT:STIME 4" !Sets the state to on and the timeout
SETup:PVTime:TIMeout:TIME
Function
Sets/queries the timeout value in seconds that will be used for power versus time measurements.
The units (S|MS) are optional, if no units are specified then units default to S.
Setting
Range: .1 to 999
Resolution: .1
Query
Range: .1 to 999
Resolution: .1
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:PVTIME:TIMEOUT:TIME 6" !Sets the timeout value to 6 seconds.
430
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pvtime.fm
"SETup:PVTime:TIME[:OFFSet]"
"FETCh:PVTime:POWer[:ALL][:MAXimum]?" on page
!value to 4 seconds.