Agilent Technologies 8960 10 Series Reference Manual page 417

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:ORFSpectrum:MODulation:FREQuency:POINts?
Function
Queries the number of frequency offsets currently on during an ORFS due to modulation
measurement. See
Query
Range: 0 to 22
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:ORFSpectrum:SWITching:COUNt[:SNUMber]
Function
Sets /queries the ORFS due to switching multi-measurement count value and turns the state on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:COUNT:SNUMBER 55" !Sets the
SETup:ORFSpectrum:SWITching:COUNt:NUMBer
Function
Sets/queries the ORFS due to switching multi-measurement count value. see
Spectrum Measurement Description" on page 75
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:COUNT:NUMBER 15" !Sets the
"Output RF Spectrum Measurement Description" on page 75
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_orfspectrum.fm
!multi-measurment
!value to 10 and the
!state to on.
"Output RF
!multi-measurement
!count number for ORFS
!due to switching
!to 15.
SETup:ORFSpectrum
417