Agilent Technologies 8960 10 Series Reference Manual page 418

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:ORFSpectrum
SETup:ORFSpectrum:SWITching:FREQuency[:OFFSet]
Function
Sets/queries the list of output RF spectrum due to switching frequency offsets. Each offset listed
in the command is turned on by default. If no frequency offsets (null list) are sent, the output RF
spectrum due to switching measurement will not be made. The units (GHZ|MHZ|KHZ|HZ) are
optional, if no units are specified than units default to HZ. See
Measurement Description" on page 75
Range: 0 to 8 comma-separated values ranging from −1.8 MHz to −10 Hz, and +10 Hz to +1.8
Setting
MHz
Resolution: 10 Hz
Range: 0 to 8 comma-separated values ranging from −1.8 MHz to −10 Hz, and +10 Hz to +1.8
Query
MHz
Resolution: 10 Hz
*RST Setting
Offset 1 = 400.0 kHz
Offset 2 = 600.0 kHz
Offsets 3 to 8 off
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET 400 KHZ, 700 KHZ"
!Turns on the first two ORFS due to switching measurement offsets and sets them
!to 400 kHz and 700 kHz offsets. All other offsets are in the off state.
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET 700 KHZ" !Turns on the
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET" !Turns all of the ORFS
418
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_orfspectrum.fm
"Output RF Spectrum
!first ORFS
!due to
!switching
!measurement
!offset and
!sets it to
!700 kHz
!offsets. All
!other offsets
!are in the
!off state.
!due to switching
!measurements offsets
!to off.