Agilent Technologies 8960 10 Series Reference Manual page 397

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:CONTinuous
February 14, 2000
SETup
"Diagram Conventions" on page 213
SETup[:ALL]:CONTinuous
Function
Sets trigger arm to OFF (single trigger) or ON (continuous trigger) for all measurements. See
"Trigger Arm (Single or Continuous) Description" on page
At power on and a (manual user) full preset the trigger arm is set to continuous. Partial preset
has no effect on the trigger arm state. See
Remote full preset sets the trigger arm to single, this is the recommended trigger arm for any
remote measurements.
Trigger arm may be set and queried for each individual measurement.
Setting
Range
• Continuous trigger = ON
• Single trigger = OFF
*RST Setting
Single
Programming Example
OUTPUT 714;"SETUP:ALL:CONTINUOUS:OFF" !Sets trigger arm for all measurements
OUTPUT 714;"SETUP:PVTIME:CONTINUOUS:OFF" !Sets trigger arm for power versus
OUTPUT 714;"SETUP:TXPOWER:CONTINUOUS?" !Queries trigger arm for TX power
:CONTinuous
[:ALL]
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_continuous.fm
"Preset Descriptions" on page
!to single.
!time measurements to single each
!measurment can be set individually.
!measurements. Trigger arm may be queried
!only one measurement at a time.
:OFF|:ON
151.
535.
SETup:CONTinuous
397