Burst Synchronization Of Measurements; Measurement Synchronization - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

Burst Synchronization of Measurements

Measurement Synchronization

Measurement Synchronization Description
Measurement synchronization determines how a measurement's time reference is determined from the
measurement data (a sampled time record). Measurement synchronization occurs after the measurement data
is captured.
For the transmit power and ORFS (switching and modulation) measurements, the RF amplitude of the input
signal is used for measurement synchronization. For Phase and Frequency Error Measurement and the Power
versus Time Measurement (see
"Power versus Time Measurement Description" on page
measurement synchronization:
• Midamble
• RF Amplitude
• None
Selecting midamble causes the test set to use the input signal's midamble data to determine the
measurement's time reference. A measurement is capable of midamble synchronization if the test set is able to
determine transmitted data from measurement samples (i.e. perform demodulation). Midamble
synchronization is not available for Transmit Power measurements, however Power vs. Time measurements
performs the average power measurement and does provide midamble synchronization.
When the test set's operating mode is "test mode" or when the cell activated state is "off", the
NOTE
burst type may need to be specified before the test set can synchronize to the input signal's
midamble. See
Selecting RF amplitude causes the test set to use the input signal's rising and falling edges (if edges are
detected within the sampled time record) to determine the measurement's time reference. If a non-bursted
signal was sampled, the measurement's time reference will be developed using the beginning and end of the
sampled time record, and the samples used for making the measurement will be taken from the middle of the
time record.
Selecting None causes the test set to perform measurements exactly as if RF amplitude was chosen.
An integrity indicator is returned for each completed measurement. Integrity errors are prioritized so that
when multiple errors occur, the highest priority error is returned first, as the root error. The integrity indicator
returns a number from 0 to 16, where zero = normal. The following integrity indicators reveal problems with
measurement synchronization:
"Phase and Frequency Error Measurement Description" on page 82
"Expected Burst" on page
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_bursr_sync.fm
88.), however, there are three possible settings for
526.
Burst Synchronization of Measurements
and
115

Hide quick links: