CALL:MS:REPorted:REVIsion?
Function
Query the MS to determine which Phase of GSM standards it complies with.
Query
range: UNKNown | PHASe1 | PHASe2
*RST Setting
PHAS2
Programming Example
OUTPUT 714;"CALL:MS:REPORTED:REVISION?"
CALL:MS:REPorted:RXLevel[:LAST]?
Function
Received level of the TCH in dB (relative to -110 dBm) which the MS measured during the last
SACCH multiframe.
Query
range: 0 to 63 and 9.91E+37
resolution: 1
*RST Setting
9.91E+37 (NAN)
Programming Example
OUTPUT 714;"CALL:MS:REPORTED:RXLEVEL:LAST?"
CALL:MS:REPorted:RXLevel:NEW?;NEW?;NEW?
Function
Queries the received level of the TCH in dB (relative to -110 dBm) which the MS measured.
Each time the :NEW? query is sent the test set hangs until report results from that
measurement period are sent.
A hanging query that will not return until the MS reports a new SACCH message to test set.
This will return 3 variables the first two must be ignored, the value from the third new query
is valid data.
Query
range: 0 to 68 and 9.91E+37
resolution: 1
*RST Setting
9.91E+37 (NAN)
Programming Example
OUTPUT 714;"CALL:MS:REPORTED:RXLEVEL:NEW?;NEW?;NEW?" !The third result is valid.