Standard Event Status Register Bit Assignment
Bit
Binary
Number
Weighting
15
32768
14
16384
13
8192
12
4096
11
2048
10
1024
9
512
8
256
7
128
6
64
5
32
4
16
3
8
Condition
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Reserved by IEEE.
Power On
Reserved for future use.
Command Error
Execution Error
Device Dependent Error
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_STATUS_STD_EVNT.fm
Description
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit will always be 0.
This bit is set to 1 if the power supply has been turned
off and on since the last time this register was read or
otherwise cleared. Defined in "IEEE Std.
488.2-1992",11.5.1.1.2
This bit will always be 0.
This bit is set to 1 if the test set detects an error while
trying to process a command. The following events
cause a command error:
• An IEEE 488.2 syntax error. The test set received
a message that did not follow the syntax defined
by the standard.
• A semantic error. For example the test set received
an incorrectly spelled command.
• The test set received a group execution trigger
(GET) inside a program message
This bit is set to 1 if the test set detects an error while
trying to execute a command. The following events
cause a execution error:
• A <PROGRAM DATA> element received in a
command is outside the legal range for the test set,
or it is inconsistent with the operation of the test
set.
• The test set could not execute a valid command
due to some test set hardware/firmware condition.
This bit is set to 1 if a test set operation does not
execute properly due to an internal condition (such as,
overrange). This bit indicates that the error was not a
command, query, or execution error.