Preset Descriptions
Description
The test set is capable of accepting several different preset commands.
At no time during a preset operation, will transmit power exceed the last user setting of the transmit power.
The input power will not be set to any value lower than the last user setting of the input power. This is to avoid
power spikes on the output and possible receiver damage on the input during transitions associated with
preset operations.
Examine the results in Tables 4 and 5 to determine which preset to use for your situation.
Partial Preset
Users save setup time with a partial preset because measurement setup parameters remain unchanged. This
is the recommended way to place the test set in a known condition.
Press the green PRESET key on the front panel to perform a partial preset.
Table 1. Partial Preset Behavior
Function
Trigger Arm
Measurement parameters
Call in progress
Operating Mode
Measurements
Measurement results
Cell Activated State
Cell Power State
Call Control Status
Call Counters
Call Error Counters
SACCH
Measurement integrity
indicator
Example
OUTPUT 714;"SYSTEM:PRESET3" !Command for a partial preset when user in
SYSTEM:PRESET3 is the recommended command for a Partial Preset operation. The SYSTEM:PRESET[1]
Result
no change
no change
aborted
Active Cell
aborted and inactivated
NAN
ON
ON
Idle
cleared
cleared
cleared
1 = no result available
!remote operation.