Sacch Report Measurement Results - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SACCH Report Measurement Descriptions
Figure 4. SACCH Report Measurement Cycle
MS reported values
averaged from 4 frames
t0
If several SACCH reported values are needed from the same report, the first value needed should be queried
three times (to receive a stable new report). Then the additional values should be immediately queried using
the :LAST? query before the next report arrives or the measurement parameters are changed again.
The :LAST? query is not a hanging query; values are returned from the last SACCH report. As shown in the
following program example (line 60), the :LAST? command is optional. If :NEW? is not used in the
MS:REPORTED command, the :LAST value is automatically reported.
Program example:
10
OUTPUT 714;"CALL:CELL:POW -83"
20
OUTPUT 714;"CALL:MS:TADV 11"
30
OUTPUT 714;"CALL:MS:TXL 11"
40
OUTPUT 714;"CALL:MS:REPORTED:TXL:NEW?;NEW?;NEW?"
50
ENTER 714;Ignore_result,Ignore_result,Valid_result ! Only use Valid_result
60
OUTPUT 714;"CALL:MS:REPORTED:RXL?;TADV?"
70
ENTER 714;Rceived_lvl,Timing_adv
80
END

SACCH Report Measurement Results

• MS TX level reported results reflect the value set in the Call Parms, MS TX Level field.
• TCH timing advance reported results reflect the value set in the Call Parms, Timing Advance field.
• RX Level reported reflects the received level of TCH in dB, from the Call Parms, Cell Power field that the
MS measured during the preceding SACCH.
• RX Qual reported reflects the perceived quality of the signal used for the RX level SACCH report.
104
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\meas_sacch_rep_desc.fm
averaged from 4 frames
t1
:NEW?
t0: measurement parameter is changed to a new value.
t1-t2: MS measures the new value.
t3: the test set receives the first SACCH report that
contains valid results reflecting the new parameter value.
MS reported values
:NEW?
! Query 3 times
! Additional values
MS reported values
averaged from 4 frames
t2
t3
:NEW?