A3 Front-End Assembly

This procedure applies to:
General Instructions
• The A3 Front-End assembly has static-sensitive components. Read
CAUTION
Chapter 15
• Hold the A3A3 Low-Pass Filter, A3A9 Bandpass Filter, and FL1
Low-Pass
their semi-rigid
are allowed to twist.
• The bracket assembly is made up of four separate brackets. Whenever
the brackets are called out in a replacement procedure, they are
designated by the letters A through D, as illustrated in
To disassemble the A3 Front-End assembly, refer to the foldouts at the
end of
the various option and model mixes.
The torque specification for all of the SMA connections within the A3
Front-End assembly is 10 inch-pounds. Be sure that the assemblies are
securely mounted during initial loosening or final torquing of the
semi-rigid cables.
Chapter 7
8592L spectrum analyzers
8593E spectrum analyzers
8594E spectrum analyzers
8594L spectrum analyzers
8594Q QAM analyzers
8595E spectrum analyzers
8596E spectrum analyzers
before proceeding.
Filter firmly in place when loosening or tightening either of
cables. The cables can be easily damaged if the filters
Chapter
12.
Chapter 11
Replacing Major Assemblies
provides cable location information for
A3 Front-End Assembly
Figure
7-6.
311