Measurement Examples

Measuring the Deviation from a Linear Phase

Measuring the Deviation from a Linear Phase
This section illustrates how to determine the deviation from a linear phase in the pass band
of a 1.09 GHz bandpass filter.

Evaluation Steps

Here, the DUT is evaluated by following the steps described in Table 14-5.
Table 14-5
Evaluating the Deviation from a Linear Phase
Step
"1. Connect the DUT" on page 302
"2. Define the Measurement
Conditions" on page 303
"3. Execute the Calibration" on
page 303
"4. Connect the DUT and Execute the
Auto Scale" on page 304
"5. Specify the Electrical Delay" on
page 304
"6. Measure the Deviation from a
Linear Phase" on page 305

1. Connect the DUT

Connect the DUT as shown in Figure 14-21.
302
Description
The DUT is connected.
The measurement conditions are defined.
The calibration is executed.
The DUT is connected again to execute the auto scale function.
The electrical delay is specified.
The statistics data function (peak-to-peak) is used to determine the
deviation from a linear phase.
Chapter 14