Editing/Generating Sensor Calibration Tables - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

5
Using 8480 Series Power Sensors

Editing/Generating Sensor Calibration Tables

To help achieve the best accuracy in your measurement you can enter the values
supplied for the sensors you are using by editing the installed sensor calibration
tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any additional
tables. However, you can edit or delete the content of each table. If you need
another table you should edit and rename one of the tables. Each calibration table
can contain a maximum of 80 frequency points.
To view the calibration tables currently stored in the power meter, press
Tables
Figure
Table 5-2
Installed power sensor models
Table
0
1
2
3
4
[a] DEFAULT is a sensor calibration table where the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter.
[b] The 8482B and 8482H power sensors use the same data as the 8482A.
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9. These tables do not contain any data when the power meter is
shipped from the factory.
166
,
Sensor Cal Tables
5-6.
Sensor Model
[a]
DEFAULT
8481A
[b]
8482A
8483A
8481D
. The Sensor Tbls screen is displayed as shown in
Table
5
6
7
8
9
Keysight N1913/1914A User's Guide
Sensor Model
8485A
R8486A
Q8486A
R8486D
8487A
,

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a

Table of Contents