Tests, Adjustments, and Troubleshooting
Test Management Environment (TME) Performance Tests
Table 3-2
Required Test Equipment (TME Performance Testing)
Test Equipment
50 Ω termination
10 dB fixed attenuator
(2required)
Calibration kit, 3.5 mm
ECal module, 3.5 mm
Cables
RF 3.5 mm (m) to 3.5 mm (f)
(3 required)
Adapters
3.5 mm (f) to 3.5 mm (f)
3.5 mm (m) to 2.4 mm (m)
(2 required)
3.5 mm (m) to Type-N (m)
BNC (m) to Type-N (f)
3.5 mm (m) to 3.5 mm (m)
(2 required)
a. N1914A power meters with serial number prior to MY53040007 require Service Note N1914A-07 which fixes a power
supply ground loop. The ground loop injects noise into the measurement circuits which can cause unstable measure-
ments at low (-30 dBm) power levels.
b. The frequency reference is optional. It may be a 10 MHz "house standard", or it may be the internal frequency refer-
ence time-base of the counter used for the frequency measurement. See the Frequency Reference Explanation section
in the Frequency Accuracy test for details.
Guidelines for Ordering Power-Sensor Calibration Service
Use the following guidelines to ensure power sensors used by the N7800A
application receive the correct calibration service:
1. Order Option 1A7 when ordering new power sensors that will be used as
2. In the Americas region, order the Keysight Cal + Measurement Uncertainty
3. Outside the Americas region, order ISO 17025/ILAC-G8 calibration service
46
Recommended Model
909D Option 301
8493C Option 010
85052B
(only the loads are used)
N4691-60004
(order as N4691B Option M0F)
8121-2111
85059B
11904A
1250-1743
1250-1477
83059A
working standards in N7800A TME calibrations.
for power sensor recalibration.
for power sensor recalibration.
Alternate Model
909D Option 40
None
85052D
N4691-60008
1250-1749
1250-1748
Keysight P937xA PXIe Service Guide