Prepare the tubes
Prepare the instrument
Check the
consumables
status
SeqStudio
Genetic Analyzer Instrument and Software User Guide
™
On a clean and level surface:
1. Place the tubes in the MicroAmp
the tubes.
2. Prepare the sample according to your application protocol, then pipet the sample
into the tubes.
3. Place a septum on the tubes.
a. Align the holes of the septa with
the tubes.
b. Press gently until the septum
are inserted into position in
each tube.
4. Centrifuge the tube assembly briefly
to collect the contents at the bottom
of each tube.
Centrifuge the tube assembly again
if the contents are not at the bottom
of the tubes.
5. Heat-denature as required.
6. Place the tube assembly on ice for 3 minutes.
Load the tube assembly onto the instrument immediately or keep the tubes on ice and
protected from light until they are loaded onto the instrument.
1. Touch
Settings4Consumable status.
2. Ensure that:
• Sufficient consumables are installed for the run.
• Installed consumables have not exceeded their expiry date.
Chapter 2 Prepare the samples and the instrument
™
96-well tray, then place the tray retainer over
1
2
3
4
1
Septum
2
96-well retainer
3
8-tube strip
MicroAmp
4
Prepare the instrument
96-well tray
™
2
39