Deprecated Commands - R&S FSW-K7 User Manual

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<Format>
<DataFormat>
Usage:
MMEMory:STORe<n>:IQ:STATe 1, <FileName>
This command writes the captured I/Q data to a file.
The file extension is *.iq.tar. By default, the contents of the file are in 32-bit floating
point format.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "Memory full" error can occur although
the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSW User Manual.
Suffix:
<n>
Parameters:
1
<FileName>
Example:
Manual operation:

11.10 Deprecated Commands

The following commands are provided for compatibility to other signal analyzers only.
For new remote control programs use the specified alternative commands.
CALCulate<n>:FEED
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
FLOat32
32-bit floating point format.
*RST:
FLOat32
COMPlex
Exports complex data.
*RST:
COMPlex
Query only
.
irrelevant
String containing the path and name of the target file.
MMEM:STOR:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Stores the captured I/Q data to the specified file.
See
"I/Q Export"
on page 164
..................................................................................................... 376