SYSTem:SEQuencer <State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (INIT:SEQ...) are executed, otherwise an error will
occur.
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S FSW User Manual.
Parameters:
<State>
Example:

11.6 Configuring the Result Display

The following remote commands are required to configure the screen display in a
remote environment.
General Window
Working with Windows in the
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
SYST:SEQ:OFF
Deactivates the scheduler
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a new data measurement and waits for the end of the
sweep.
INIT:SEQ:REFR
Refreshes the display for all channels.
Event
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands (INIT:
SEQ...) are not available.
*RST:
0
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF