Parameters:
<State>
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
Usage:
Manual operation:
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S FSW or limiting the dynamic range by an
S/N ratio that is too small.
Example:
Usage:
Manual operation:
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous] <State>
Activates automatic scaling of the y-axis in all diagrams according to the current mea-
surement results. Currently auto-scaling is only available for AF measurements. RF
power and RF spectrum measurements are not affected by the auto-scaling.
Parameters:
<State>
Example:
Manual operation:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
ON | 1
The measurement for automatic adjustment waits for the trigger.
OFF | 0
The measurement for automatic adjustment is performed imme-
diately, without waiting for a trigger.
*RST:
1
ADJ:FREQ
Event
See
"Adjusting the Center Frequency Automatically (Auto Freq)"
on page 130
ADJ:LEV
Event
See
"Setting the Reference Level Automatically (Auto Level)"
on page 89
ON | OFF | 1 | 0
*RST:
0
SENS:ADJ:SCAL:Y:AUTO ON
See
"AF Auto Scale"
Configuring the Measurement
on page 122
286

Advertisement

Table of Contents
loading

Table of Contents