Figure 4-6: Test setup for reflection measurement
Generated signal input
In order to use the functions of the external generator, an appropriate generator must
be connected and configured correctly. In particular, the generator output must be con-
nected to the RF input of the R&S FSW.
External reference frequency
In order to enhance measurement accuracy, a common reference frequency should be
used for both the R&S FSW and the generator. If no independent 10 MHz reference
frequency is available, it is recommended that you connect the reference output of the
generator with the reference input of the R&S FSW and that you enable usage of the
external reference on the R&S FSW via "SETUP" > "Reference" > "External Refer-
ence".
For more information on external references see the "Instrument Setup" section in the
R&S FSW User Manual.
Connection errors
If no external generator is connected, if the connection address is not correct, or the
generator is not ready for operation, an error message is displayed (e.g."Ext. Genera-
tor TCPIP Handshake Error!", see
on page 43).
4.7.4.2
Overview of Supported Generators
Generator type
SGS100A6
SGS100A12
SGT100A3
SGT100A6
1) Requires firmware version V2.10.x or higher on the signal generator
2) Requires firmware version V1.10.x or higher on the signal generator
3) Requires the option SMR-B11 on the signal generator
4) Requires firmware version V3.20.200 or higher on the signal generator
User Manual 1173.9240.02 ─ 23
Receiving Data Input and Providing Data Output
Chapter 4.7.4.8, "Displayed Information and
TTL support
Generator type
-
SMP02
-
SMP03
-
SMP04
-
SMP22
Measurement Basics
Errors",
TTL support
X
X
X
X
35

Advertisement

Table of Contents
loading

Table of Contents