Parameters:
<FastExport>
SYSTem:COMMunicate:RDEVice:OSCilloscope[:STATe] <State>
Activates the optional 2 GHz bandwidth extension (R&S FSW-B2000).
Note: Manual operation on the connected oscilloscope, or remote operation other than
by the R&S FSW, is not possible while the B2000 option is active.
Parameters:
<State>
Example:
Manual operation:
SYSTem:COMMunicate:RDEVice:OSCilloscope:ALIGnment:STEP[:STATe]?
Performs the alignment of the oscilloscope itself and the oscilloscope ADC for the
optional 2 GHz bandwidth extension (R&S FSW-B2000). The correction data for the
oscilloscope (including the connection cable between the R&S FSW and the oscillo-
scope) is recorded. As a result, the state of the alignment is returned.
Alignment is required only once after setup. If alignment was performed successfully,
the alignment data is stored on the oscilloscope.
Thus, alignment need only be repeated if one of the following applies:
A new oscilloscope is connected to the IF OUT 2 GHZ connector of the R&S FSW
A new cable is used between the IF OUT 2 GHZ connector of the R&S FSW and
the oscilloscope
A new firmware is installed on the oscilloscope
Return values:
<State>
Example:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
ON | OFF | 1 | 0
ON | 1: Disables the display update for maximum export speed.
OFF | 0: Enables the display update. The export is slower.
*RST:
1
ON | OFF | 1 | 0
ON | 1
Option is active.
OFF | 0
Option is disabled.
*RST:
0
SYST:COMM:RDEV:OSC ON
See
"B2000 State"
on page 79
Returns the state of the second alignment step.
ON | 1
Alignment was successful.
OFF | 0
Alignment was not yet performed (successfully).
SYST:COMM:RDEV:OSC:ALIG:STEP?
//Result: 1
Configuring the Measurement
206