R&S FSW-K7 User Manual page 308

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
[SENSe:]ADEMod<n>:AM:RELative:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod<n>:FM[:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod<n>:FM:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod<n>:PM[:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod<n>:PM:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod<n>:SPECtrum:RESult? <TraceMode>
This command reads the result data of the evaluated signal in the specified trace
mode. The data format of the output data block is defined by the FORMat command
(see
The trace results are configured for a specific evaluation. The following table indicates
which command syntax refers to which evaluation method, as well as the output unit of
the results.
Command syntax
AM[:ABSolute][:TDOMain]
AM:RELative[:TDOMain]
AM:RELative:AFSPectrum
FM[:TDOMain]
FM:AFSPectrum
PM[:TDOMain]
PM:AFSPectrum
SPECtrum
Suffix:
<n>
Query parameters:
<TraceMode>
Example:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
on page 309).
FORMat[:DATA]
.
irrelevant
WRITe | AVERage | MAXHold | MINHold | VIEW
The specified trace mode must be one of those configured by
SENS:ADEM:<Evaluation>:TYPE, see
]ADEMod<n>:SPECtrum[:TYPE]
query error is generated.
ADEM:AM AVER,MAXH,MINH
Sets up RF time domain results to be measured
INIT; *WAI
Starts measurement and waits for sync
FORM ASC
Selects output format
ADEM:AM:RES? AVER
Reads RF time domain average results
ADEM:AM:RES? MAXH
Reads RF time domain max hold results
ADEM:AM:RES? MINH
Reads RF time domain min hold results
Evaluation method
RF time domain
AM time domain
AM spectrum
FM time domain
FM spectrum
PM time domain
PM spectrum
RF spectrum
Retrieving Results
Output unit
dBm
%
%
kHz
kHz
rad or °
rad or °
dBm (logarithmic display)
or V (linear display).
[SENSe: