R&S FSW-K7 User Manual page 362

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Suffix:
<n>
<k>
Parameters:
<Mode>
Manual operation:
CALCulate<n>:LIMit<k>:UPPer:OFFSet <Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
<k>
Parameters:
<Offset>
Manual operation:
CALCulate<n>:LIMit<k>:UPPer:SHIFt <Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
<k>
Parameters:
<Distance>
Usage:
Manual operation:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
Window
Limit line
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
See
"X-Axis"
on page 159
.
irrelevant
Limit line
Numeric value.
*RST:
0
Default unit: dB
See
"Y-Offset"
on page 157
.
irrelevant
Limit line
Defines the distance that the limit line moves.
The unit depends on
on page 360.
Event
See
"Shift y"
on page 160
CALCulate<n>:LIMit<k>:UNIT
Analyzing Results
362

Advertisement

Table of Contents
loading

Table of Contents