R&S FSW-K7 User Manual page 226

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<State>
Example:
Manual operation:
[SENSe:]PMETer<p>:UPDate[:STATe] <State>
This command turns continuous update of power sensor measurements on and off.
If on, the results are update even if a single sweep is complete.
Suffix:
<p>
Parameters:
<State>
Example:
Manual operation:
UNIT<n>:PMETer<p>:POWer <Unit>
This command selects the unit for absolute power sensor measurements.
Suffix:
<n>
<p>
Parameters:
<Unit>
Example:
Manual operation:
UNIT<n>:PMETer<p>:POWer:RATio <Unit>
This command selects the unit for relative power sensor measurements.
Suffix:
<n>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
ON | OFF | 1 | 0
*RST:
0
PMET1 ON
Switches the power sensor measurements on.
See
"State"
on page 84
See
"Select"
on page 84
.
1...4
Power sensor index
ON | OFF | 1 | 0
*RST:
0
PMET1:UPD ON
The data from power sensor 1 is updated continuously.
See
"Continuous Value Update"
.
irrelevant
1...4
Power sensor index
DBM | WATT | W
*RST:
DBM
UNIT:PMET:POW DBM
See
"Unit/Scale"
on page 85
.
irrelevant
Configuring the Measurement
on page 84
226