[SENSe:]CORRection:CVL:PORTs <PortNo>
This command defines the mixer type in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<PortType>
Example:
Manual operation:
[SENSe:]CORRection:CVL:SELect <FileName>
This command selects the conversion loss table with the specified file name. If
<file_name> is not available, a new conversion loss table is created.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<FileName>
Example:
Manual operation:
[SENSe:]CORRection:CVL:SNUMber <SerialNo>
This command defines the serial number of the mixer for which the conversion loss
table is to be used. This setting is checked against the current mixer setting before the
table can be assigned to the range.
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
string
Name of mixer with a maximum of 16 characters
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX 'FS_Z60'
See
"Mixer Name"
[SENSe:]CORRection:CVL:SELect
2 | 3
*RST:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
See
"Mixer Type"
String containing the path and name of the file.
CORR:CVL:SEL 'LOSS_TAB_4'
See
"New Table"
See
"Edit Table"
See
"File Name"
Configuring the Measurement
on page 64
on page 202).