R&S FSW-K7 User Manual page 113

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
The "Time per Division" softkey is available from the main "Analog Demodulation"
menu.
Zero Phase Reference Position (PM Time Domain only)
Defines the position at which the phase of the PM-demodulated signal is set to 0 rad.
The entry is made with respect to time. In the default setting, the first measured value
is set to 0 rad.
This setting is only available for PM time domain displays with DC coupling.
Remote command:
[SENSe:]ADEMod<n>:PM:RPOint[:X]
Phase Wrap On/Off (PM Time Domain only)
Activates/deactivates the phase wrap.
On
Off
This setting is only available for PM time domain displays with DC coupling.
5.7.2 Demodulation Spectrum
Access: "Overview" > "Demod Settings" > "Spectrum"
Or: "Meas Setup" > "Demod" > "Spectrum" tab
The demodulation spectrum defines which span of the demodulated data is evaluated.
Depending on the evaluation (AF or RF display), the settings vary.
AF
RF
5.7.2.1
AF Evaluation
Access: "Overview" > "Demod Settings" > "Spectrum"
Or: "Meas Setup" > "Demod" > "Spectrum" tab
These settings are only available for AF Spectrum evaluations, not in the time domain.
User Manual 1173.9240.02 ─ 23
The phase is displayed in the range ±180° (± Π). For example, if the phase exceeds +180°,
360° is subtracted from the phase value, with the display thus showing >-180°.
The phase is not wrapped.
Evaluation........................................................................................................113
Evaluation........................................................................................................115
on page 268
Configuration
Demodulation
113